Measuring procedure
- Switch-on the spectrometer (see picture step
1).
- Switch-on one of the three light emitters -
blue, green or red - of a tri-colour light emitting
diode (LED) (see picture step 2).
- Set the light intensity of the selected light
source - blue, green or red -to the maximal
by turning the appropriate knob to the far right
position (see picture step 3).

- Insert a blister into the measuring chamber,
so that the hollow containing a blank occupies the
measuring site (see picture step 4).
- Set the transmittance to 100.0 %. Use knobs
for rough and fine regulation of the transmittance
(see picture steps 5 and 6).
- Change the position of the blister in the measuring
chamber, so that a calibration or examined solution
occupies the measuring site (see picture step 4).
- Read a transmittance from the display (see picture
step 7).
- Calculate absorbance from the measurement of
transmittance (T) as given below:
A = - log (T/100)
Note. Within the same experiment volumes of all solutions
in hollows of a blister must be equal.