PROTOCOL N° 1
WATER pH MEASUREMENT
EQUIPMENT AND REAGENTS
- pH meter with electrodes (glass electrode plus reference electrode,
or combined glass/reference electrode)
- Buffer solutions: pH=4, pH=7 and pH=10
PROCEDURE
- Take approximately 30 mL of the water sample and place it in a beaker,
- After calibration of measuring set-up rinse the electrode(s)
with distilled water, gently dry off the excess
water with tissue paper and insert in the water sample allowing
1 min for equilibration,
- Record temperature and pH values.
|
|
|