introduction

calcium

hydrogen phosphate

chloride

ammonium

nitrite

hydrogen carbonate

total water hardness

pH

conductivity and TDS

8. A POTENTIOMETRIC DETERMINATION OF THE
pH OF A WATER SAMPLE

8.1 Objective

Use a two-point calibration and determine the pH of a water sample. Register the asymmetric potential and the slope of the pH electrode.

8.2 Background

The measuring system for a potentiometric determination of pH consists of a mV-meter with a high input resistance (> 1 TΩ), a combined pH electrode with a glass membrane, a temperature sensor and the examined solution. In the combined electrode, a glass electrode and a reference electrode of the Ag-AgCl type are assembled in the same body.

The relation between the electrode potential and the activity of the H3O+ ions in a solution is defined by the Nernst equation. For practical use the relation can be written as follows:

E = k + s·log[H3O+]      or     E = k – s·pH

where the letter s stands for the slope. The theoretical slope at standard temperature is 0.0592 V. The pH-meter is built so that at a pH of 7.00 the potential should be 0.0 mV. However, real electrodes seldom perform ideally, and as a consequence the pH-meter should be calibrated before use. At least a two-point calibration should be used; this involves two calibration buffers for which the pH is reliably known at the working temperature. The calibration process provides an insight into the asymmetric potential and the real slope, or the real relative slope, of the electrode in use. The asymmetric potential provides information about how much the potential at pH 7.00 differs from the expected zero potential.   

        The calibration ensures the reliability of the pH determination of the examined solution. However, it is important that the pH of examined solutions falls within the range of the calibration and that the temperature is very close to that at which the calibration was performed.

8.3 Equipment and reagents

  • A WTW pH 323/325 portable pH-meter.
  • A calibration buffer of pH 4.
  • A calibration buffer of pH 10.

8.4 Procedure

Calibration of the pH-meter*

The electrode glass membrane is very sensitive and can break easily. Use a polymeric beaker as a drain while rinsing the electrode. Rinse the electrode membrane with deionised water between measurements. Wipe the membrane gently with tissue paper.

Use the pH 4 and pH 10 calibration buffers for the calibration procedure. Mix the buffer solution thoroughly before use. Immerse the electrode in the first buffer solution.

Repeatedly press the CAL button until the ASY calibration mode appears on the screen of the pH-meter. Next, press the RUN button. Wait until the pH value stabilises. Use the appropriate arrow button and set the correct pH for the selected buffer at the working temperature. Confirm the setting with the RUN button. Register the asymmetric potential and the temperature.

Rinse and wipe the surface of the electrode membrane and then immerse the electrode in another buffer solution. Repeatedly press the RUN button until it is possible to input the pH again. Wait until the pH value stabilises and then set the correct pH value as before. Confirm with the RUN button. Register the slope of the electrode. Rinse and wipe the electrode membrane.

* For other instruments follow the instructions of the producer.
 

Determining the pH of the water sample*

Immerse the clean and dry electrode membrane into the sample solution. Select pH measurement by pressing the button pH/mV, and then confirm the selection with the RUN button. Wait until the pH stabilises. Write the pH of the water sample and the other requested data in the table below.  

* For other instruments follow the instructions of the producer.

?

Asymmetric potential

Slope

pH

 

 

 

? Also write the pH in the table “Basic data on water samples”, under the serial number of the sample.

The guide value for pH in drinking water which is set by the EU Drinking Water Directive (Council Directive 80/778/EEC) is between 6.5 and 8.5.